On we started the Liquefier A after cleaning up the Dewar A level sensors using the procedure (docdb). According to the procedure, the bottom tube did not vent gHe on when the cold head temperatures are slightly above 240 K. So, we had to wait until all the cold heads are above 272K.
ImageImage Added


Since then (03/06/2023), Dewar A wasn't producing liquid until the 03/19/2023 afternoon. Below is the log of things that we did.


Around 11:26am on 03/12/2023
1. Set the "Mode End: Coldhead Temperature" to 3 K
    Screen_Shot_2023-03-12_at_11.26.33_AM.pngImage Added
2. Had to switch the LQ-A to the Manual mode for a minute and the switched back to Liquefy mode to enable the change made in the previous step.
3. See the cold-head temperatures since 03/06/2023.
    Screen_Shot_2023-03-12_at_9.21.51_PM.pngImage Added
4. Flow change after making the change on Mode End: Coldhead Temperature" to 3 K.
    Screen_Shot_2023-03-12_at_9.22.28_PM.pngImage Added