Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

The issue was that, the FCAI hasn't go up beyond ~8slm although FCAQ is set to 24slm.
Then, on 03/14/2023 around 4:20pm CST, we closed the NV501A needle valve as Adam suggested, to improve the Liquefaction on LQ-A. It was between 1/4 and 1/2 a turn opened position before it was completely closed.
     Initial position                                                                                 Completely closed position
NV501_03-14-2023_Before_02.jpgImage Added                        NV501_03-14-2023_After_02.jpg



On 03/15/2023 around 2:20 pm
We opened the NV501A needle valve by two full turns from the current setting(fully closed position) as Joey (QT) recommended.

On 03/16/2023
1.  In addition to the 2 turns (open) on 03/15/2023, NV501A was opened with one more full turn.
2. We adjusted the PRV501A, in order to obtain a smooth inlet flow.
       (Initial Position)                                                                                                                                                                                                                                 (Final Position)
    IMG_2790_Before_At_16_18.jpgIMG_2791_Position2_At_16_21.jpgIMG_2792_Position3_At_16_30.jpgIMG_2794_Position4_At_16_31.jpgIMG_2799_Position5_At_17_10.jpgIMG_2806_After_At_17_36.jpg

...

On 03/19/2023
After 13 days of starting the Cooldown of the Liquefier A, today the level sensor indicated that the Dewar A has started accumulating liquid. Notice that the outlet flow actual value is around ~13slm although FCAQ is still set to 24slm.



On 03/20/203

In the morning we noticed that the Liquefier A switched to the "Liquefy" mode automatically because the condition "Mode end: Dewar level" 5% was met.
Image Added
Image Added

The liquid level is ~24L and artificial blockage appeared. So, the Dewar pressure and heater were adjusted to get rid of the blockage.
  (Before)                                                                                          (after)
Image AddedImage Added







A screen video of Liquefier A

...

It seems that the green-handled regulator on the QT manifold should not be open too much.  This will make the MFC letting in the helium to the cold heads fluctuated a lot which is not optimal for cooling.  During cooldown mode you want a large flow running in and out of the dewar with the goal of around 60 slm going in and 24 slm going out.  In order for this to work the pressure in the cold heads much stay slightly above the dewar pressure so flow continuously runs to the dewar to cool things down.  The black-handled needle valve should be open a lot (maybe all the way) to keep the flow through this area as high as possible.  This will lead to lots of frost and ice build up on the dewar neck.  This should not be a problem but should be checked and defrosted regularly.  When everything is set appropriately the accumulation of liquid should happen in about 24 hours.
 

  Cooldown Mode Settings

  1. Open the NV501 (A/B) as much as it doesn't affect producing a lot of ice on the Dewar neck. This will allow a higher flow circulation between the Dewar and coldheads.
  2.  In the  "Cooldown Mode Settings": Set the outlet flow to 24slm and End mode cold head temperature to 3K
  3.  Reduce PRV501 (A/B) (green-hand valve) to barely provide the flow that FC501(A/B) flow controller asking for.

Liqefy Mode Settings
1.  Close the NV501 (A/B) until there is no ice accumulation on Dewar neck




Questions

a.) Is there a downside to having the needle valve open too much during the dewar cooldown?

...

c.) Do we understand why the overfilling (just up to 220L) leads to this contamination issue?  Can this be fixed with different tubing?
     With past experiences, the level sensor's large fluctuations happened if the Dewar was kept full ( > 210 L) for more than two or three days.

d.) How to adjust the outlet flow while in the Liquefy mode? (without going to the manual mode)
     Especially, when we want to bring down the Dewar pressure for a fill (like from 5pi to 3psi), it usually takes about 5 hrs. How can we improve that?
     By experience, we noticed in the past, if we use higher flow using the bypass SVVA/B, those solenoid valves got stuck-open, and we will have to avoid that situation too.

e.) Why Dewar pressure, inlet/outlet flows start behaving like the following when it was switched to the Idle mode (either by automatically or manually)?
     Image AddedImage AddedImage Added

e.) Purifier is still in Manual mode since the beginning, any plans to enable the "auto" mode?
     We noticed that once the Liquefiers are accidentally go to "Stop" mode, the v9 and v21 solenoid valves get closed.

f.) Calibrations of
    (i) FMR : This was done by Sam by applying a conversion equation with the characteristic parameters for gHe.
    (ii) Dewar Level sensors: We can adjust this by comparing with the manual gauges

g.) PT8:
     Image previewImage Added
     It seems that this pressure sensor is accurate at high pressure but has about ~15psi off when it's below 50 psi.
     Sam recommended this product (the same one):  https://www.automationdirect.com/adc/shopping/catalog/process_control_-a-_measurement/pressure_sensors/pressure_transmitters/sptd25-20-0300h