The fridge, magnet, and liquid nitrogen shield are hung in the "target can" which is held at vacuum by a di pump. The liquid nitrogen shield, seen in the diagram in green, protects the liquid helium components from heat radiation from the room temperature can.
The cryogenic Evaporation Fridge is designed to provide high cooling
power (1 W) at 1 K.
...