Tasks are mentioned below:

Monitor the fridge temperatures, IVC, flows (fridge, magnet, separator), magnet LHe level, etc.. the full system in general. 

In order to look out for these parameters, use the following links:

Shift Schedule

As of 2022-12-22.

Time06:00-10:0010:00-14:0014:00-18:0018:00-22:0022:00-02:0002:00-06:00
Monitoring PersonIsharaVibodhaErnestoFarooqKenichiZulkaida
On-Call PersonVibodha and Farooq

If anybody is going to shuffle the shift blocks, kindly inform the discord group in advance. 

Discord Passive Alarms

The Cryo Control Panel sends out a Discord message in case any of the following parameters become out of its valid range.  The table below was updated manually on 2022-12-23.

ParameterVILOHIUnit
IVC PressureTPG361n/a1e-5Torr
Nose LHe LevelFridge Valve6585%
Tank B, Coil T, Coil BITM10n/a5K
M1, M2, M3Roots-Pump LCW0.3n/aGPM
WFS1Roots-Pump LCW1.3n/aGPM

Non-Monitoring Tasks

The following tasks have to be done by onsite people.

As of 2022-12-22.

  1. Refill LHe to the fridge (up to Heat-Ex Top) and start the standby mode — At 06:00.
  2. Refill LHe to the fridge (up to level = 80%) — At 18:00.
  3. Refill Magnet space with LHe — At 19:00
  4. Refill LN2 to the Purifier & the shield — When the LN2 level becomes reasonably low.